Simple Chinese Operation Manual for Abend Centrifuge 5810R and 5804R

Ai Bende's large-capacity centrifuge has been approved by domestic users, but the English manual is very inconvenient. I printed a simple manual for my friends, and many users have posted it on the side of the machine. So that students can operate normally.

Eppendorf centrifuge 5810R steps:

1. Connect the power and turn on the switch next to the instrument.

2. According to the size of the centrifuge tube containing the sample, place the rotor, tighten it clockwise with a hex wrench, balance the centrifuge tube, cover the rotor cover (the cover of the large rotor should be tightened), and close the centrifuge properly. The lid makes the light on the "Open" button brighter.

3. The parameters of the centrifugation are adjusted by the Speed ​​key, the Time key, the Temp key, and the ▲, ▼ keys. Taking the Speed ​​setting as an example, press the Speed ​​button to make the speed value flash, then use the ▲, ▼ keys to change the value; if you want to set the centrifugal force (rcf), press the Speed ​​button repeatedly until the centrifugal force symbol (*) appears at the speed. The left side of the value allows you to make settings.

4. After the parameters are set, press the Start button to start the running program. During the running, the screen displays the current speed or centrifugal force, sample temperature and remaining time.

5. After the end of the operation or when the Stop button is pressed to terminate the operation, the rotor gradually decelerates until it stops. When the indicator on the Open button lights up, press the Open button to open the cover.

Note on the use of eppendorf centrifuge 5810R:

1. The centrifuge must be leveled. When the different types of tubes are centrifuged, they should be weighed with a balance. After the weight is equal, the centrifuge can be placed symmetrically.

2. For organic and corrosive samples, check for leaks and clean the inside of the centrifuge after centrifugation.

3. When using the refrigerated centrifugation operation, after the operation is completed, the condensed water in the inner chamber must be dried and the condensed water in the plate under the machine must be drained. To avoid the formation of condensed water, the cover must be opened after use.

4. When freezing without centrifugation is required, adjust the centrifugation temperature to room temperature.

5. Take your own items in time, pay attention to electricity safety, keep the laboratory clean, and not waste. When the instrument is not normal, report it in time and do not handle it by yourself.