The similarities and differences between graphite furnace atomic absorption spectrometer and flame atomic absorption spectrometer are as follows:
Graphite furnace atomizer: It is a kind of system that places the sample on the graphite tube wall, graphite platform, carbon rod sample hole or graphite crucible and electrically heated to high temperature to achieve atomization. Among them, tubular graphite furnace is the most commonly used atomizer.
The atomization program is divided into drying, ashing, atomization, and high temperature purification.
High atomization efficiency: 100% sample utilization at adjustable high temperatures.
High sensitivity: its detection limit is 10-6~10-14.
Less sample: suitable for the determination of refractory elements.
The choice of the optimum operating conditions for the graphite furnace (optimal atomization temperature of the inert gas).