TCI Syringe Pump (THR-SP200)
THR-SP200 TCI Â Syringe Pump
Features:
--Built-in multiple drugs,real-time display of target concentration curve
--Predict patient' wake-up time,convenient for clincal care
--RS232Â (Optional)
--Automatically compensate for missing dose when changing medication
--Compatible for 5,10,20,30,50(60)ml syringe
Specifications
THR-SP200 TCI syringe pumpÂ
Applicable syringes
Various brands of sterile syringes: 5ml,10ml,20ml,30ml,50ml/60ml
Infusion rate
5ml: 0.1 ml/h -150ml/hÂ
10ml: 0.1 ml/h -300ml/h Â
20ml: 0.1 ml/h -600ml/h   Â
30ml: 0.1 ml/h -900ml/h
50ml: 0.1 ml/h -1500ml/h
Bolus rate
5ml:Â 100-150ml/hÂ
10ml:Â 100-300ml/hÂ
20ml:Â 100-600ml/hÂ
30ml:Â 100-900ml/hÂ
50ml:Â 100-1500ml/hÂ
Rinse rate
Â
5ml:Â 150ml/h
10ml:Â 300ml/h
20ml: 600ml/h  Â
30ml:900ml/h
50ml:Â 1500ml/h
Infusion mode
Plasma TCI,Effect TCI,Intermittent,TIVA Mode,weight mode,
Time mode,Rate mode
Volume to be
infused (VTBI
0.1ml~1000ml
Accuracy
±2% (after syringe correct calibration)
KVOÂ Rate
0.1Â ~Â 2Â ml/h
Occlusion pressure
40.0-160.0kpa
3 adjustable occlusion pressure settings: low, middle and high
Alarms
Infusion completion, Empty, Occlusion, Near completion, Low battery,
Syringe disengaged, No AC power supply, Battery exhausted, Malfunction etc.
Battery
Rechargeable Li polymer battery,7.4V, 1900mAh
Can operate more than 6 hours after fully charged at 5ml/h., temperature 25ºC
Battery recharge time: 10h with power on, 3h with power off
Power supply
AC100~240V,50/60Hz
DC 12V ±1V
Power consumption
25VA
Waterproof Level
Class I, type CF IPx3
Dimensions
300(L)x 130(H)x 125 (W)mm, not including pole clamp
Net weight
≤2 kg
Operating conditions
Temperature: 5ºC~40ºC,
Relative humidity: 10~95%,
Atmosphere pressure: 86~106kPa
Welcome to send us your inquiry.